Spectrophotometer

Atomic Absorption Spectrophotometer BK-AA180B

Model BK-AA180B Petrol

BK-AA180B features an 8-lamp turret design, a flame atomizer, 6-level spectral bandwidth adjustment, and deuterium lamp background correction. It is suitable for the detection of over 30 flame-analysis elements, with a quantitative concentration reaching the mg/L(10⁻⁶) level and a detection concentration as low as μg/L(10⁻⁹). It is widely used in industries such as geology and mining, metallurgy, materials science, agriculture, forestry, soil and fertilizer, petroleum and chemical engineering, as well as third-party testing.

Net weight
70kg
Dimensions
1080*480*560mm
A tailored quote, without the guesswork.

Ask about availability, configuration and delivery to your location. We reply personally.

Request a quote
Enquiry only. No payment is taken.
  • Specifications confirmed with you before any order
  • International delivery arranged to your facility
  • Personal sourcing support from first message to arrival
A closer look

About this equipment

Features:

1. The instrument features a 20mm thick cast aluminum integrated optical platform with a suspended design, delivering exceptional stability of better than 0.002Abs/30min.

2. The flame atomizer system utilizes aerospace-grade polyphenylene sulfide(PPS) material with a molded one-piece construction, offering exceptional flame resistance, corrosion resistance, and high-temperature durability for extended service life.

3. The industrial-grade TA2 pure titanium burner, precision-finished with slow wire-cutting technology, provides excellent corrosion resistance and supports interchangeable wide-slit designs for high-salt sample analysis.

4. The next-generation flame control system integrates continuously adjustable gas flow, real-time status indicators, dual safety interlocks, and automatic ignition for reliable operation in demanding environments.

5. The optional oxygen-enriched flame upgrade achieves temperatures exceeding 2700°C without nitrous oxide, enabling precise determination of high-temperature elements like Ca, Al, Ba, Mo, Ti, and V.

6. Network port communication mode, offering significantly improved stability and data transfer rates compared to legacy interfaces like RS-232 and RS-485.

Technical Parameters:

Model BK-AA180B
Wavelength Range 170~900nm
Spectral Bandwidth 0.1, 0.2, 0.4, 0.8, 1.6, 2.4nm
Wavelength Accuracy ±0.1nm
Wavelength Repeatability ≤0.05nm
Detector Photomultiplier tube detector
Baseline Stability Drift ≤0.002Abs, noise ≤0.001Abs(Steady 30min)

Drift ≤0.002Abs, noise ≤0.001Abs(Dynamic 15min)

Resolution Spectral bandwidth deviation ≤0.02nm

Manganese double line valley to peak energy ratio ≤25%

Characteristic Concentration Flame method for Cu: ≤0.025μg/mL/1%
Detection Limit Flame method for Cu: ≤0.002μg/mL
Preciseness Flame method for Cu: ≤0.3%
Background Correction Ability Deuterium lamp calibration 1.0Abs, background ≥90 times
Diffraction Grating Groove 1800lines/mm, blaze wavelength 250nm
Lamp Stand 8(Support simultaneous preheating of 1~4 lights)
Standard Accessory Element lamp(Cu*1, Hg*1), software working station
Optional Accessory Oxygen-enriched high-temperature flame, wide slit combustion head, adjustable nebulizer, hydride vapor generator, audit trailing software, oil-free air compressor, computer, printer
Power Supply AC220V, 50/60Hz; 110V, 60Hz(Standard)
External Size(W*D*H) 1080*480*560mm
Package Size(W*D*H) 1220*670*840mm
Net Weight 70kg
Gross Weight 140kg

Flame System:

Air-acetylene Burner 100mm all-titanium burner
Atomization Chamber System Integrally molded from polyphenylene sulfide
Burner System Flame height is adjustable and lockable, with a 360° rotatable combustion seam
Gas Flow Control System Automatic ignition/fame-out
Reading Mode Foot pedal operation
Safety System Emergency shutdown protection switch, automatic gas cut-off/shutdown/alarm in case of abnormal status, abnormal air pressure, ignition failure, gas leakage, abnormal flow rate, etc.

Software Working Station:

Smart Optimization One-key operation for instrument optimization, supporting multi-task analysis
Repeat Measurement 1~99 replicates with auto-calculation of mean/SD/RSD
Print/Export Reports Word/Excel supported
Auto-Calibration Curve fitting, re-slope&concentration calculation, incl. standard addition method
The details

Specifications

Model
BK-AA180B
Wavelength Range
170~900nm
Spectral Bandwidth
0.1, 0.2, 0.4, 0.8, 1.6, 2.4nm
Wavelength Accuracy
±0.1nm
Wavelength Repeatability
≤0.05nm
Detector
Photomultiplier tube detector
Baseline Stability
Drift ≤0.002Abs, noise ≤0.001Abs(Steady 30min)Drift ≤0.002Abs, noise ≤0.001Abs(Dynamic 15min)
Resolution
Spectral bandwidth deviation ≤0.02nmManganese double line valley to peak energy ratio ≤25%
Characteristic Concentration
Flame method for Cu: ≤0.025μg/mL/1%
Detection Limit
Flame method for Cu: ≤0.002μg/mL
Preciseness
Flame method for Cu: ≤0.3%
Background Correction Ability
Deuterium lamp calibration 1.0Abs, background ≥90 times
Diffraction Grating
Groove 1800lines/mm, blaze wavelength 250nm
Lamp Stand
8(Support simultaneous preheating of 1~4 lights)
Standard Accessory
Element lamp(Cu*1, Hg*1), software working station
Optional Accessory
Oxygen-enriched high-temperature flame, wide slit combustion head, adjustable nebulizer, hydride vapor generator, audit trailing software, oil-free air compressor, computer, printer
Power Supply
AC220V, 50/60Hz; 110V, 60Hz(Standard)
External Size(W*D*H)
1080*480*560mm
Package Size(W*D*H)
1220*670*840mm
Net Weight
70kg
Gross Weight
140kg
Air-acetylene Burner
100mm all-titanium burner
Atomization Chamber System
Integrally molded from polyphenylene sulfide
Burner System
Flame height is adjustable and lockable, with a 360° rotatable combustion seam
Gas Flow Control System
Automatic ignition/fame-out
Reading Mode
Foot pedal operation
Safety System
Emergency shutdown protection switch, automatic gas cut-off/shutdown/alarm in case of abnormal status, abnormal air pressure, ignition failure, gas leakage, abnormal flow rate, etc.
Smart Optimization
One-key operation for instrument optimization, supporting multi-task analysis
Repeat Measurement
1~99 replicates with auto-calculation of mean/SD/RSD
Print/Export Reports
Word/Excel supported
Auto-Calibration
Curve fitting, re-slope&concentration calculation, incl. standard addition method
How ordering works

From enquiry to delivery

  1. 01Send your enquiry

    Request a quote on WhatsApp, or add several items to your enquiry list and send them together.

  2. 02Confirm the details

    We confirm configuration, voltage, accessories, availability and a delivered price.

  3. 03We source and ship

    We handle procurement, packing and international shipping to your facility.

Need a different configuration?

Tell us the application, capacity or budget. We will suggest the right model or source an alternative.

Talk to our team
Request a quote