Model
|
BK-310
|
|
Overall Performance
|
Throughput
|
300 tests/hour, ISE(Optional)
|
|
Analysis Method
|
End-point, Fixed-time, Rate(Kinetic), Turbidimetry
|
|
Certificates
|
CE, FDA, ISO9001, ISO14001, ISO13485
|
|
System
|
Optional open or close system, continuous working for 24 hours
|
|
Sample&Reagent System
|
Sample Positions
|
50 sample positions
|
|
Reagent Position
|
56 reagent positions
|
|
Sample Volume
|
2~70μl
|
|
Reagent Volume
|
20~350μl
|
|
Probe
|
With anti-collision function, liquid level detection function
|
|
Probe Washing
|
Automatic washing interior and exterior
|
|
Reagent Cooling
|
2~12℃ for 24 hours
|
|
Reaction System
|
Temperature Control
|
37±0.2℃, real-time monitoring
|
|
Cuvettes
|
120 reusable cuvettes
|
|
Mixer Probe
|
Independent stirring, automatic frequency conversion
|
|
Washing
|
Automatic cuvettes washing
|
|
Optical System
|
Light Source
|
Halogen lamp, water-cooling
|
|
Spectrophotometry
|
Post-spectral spectrophotometry
|
|
Wavelength
|
340~800nm
|
|
Absorbance
|
0~3.5Abs
|
|
Calibration&QC
|
Calibration
|
Linear: K factor, 1-point, 2-point and multipoint linear
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P
|
|
Quality Control
|
Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D)
|
|
Data Management
|
Software
|
Windows 7/8/10, 32 or 64 bit
|
|
LIS System
|
Bi-direction, support HL7 protocol
|
|
Interface
|
LAN port access
|
|
Printer
|
External, multiple reporting mode available
|
|
Working Conditions
|
Power Supply
|
AC220V±10%, 50/60Hz(Standard)
110V±10%, 60Hz(Optional), 350VA
|
|
Temperature
|
15~30℃(±2℃)
|
|
Water Consumption
|
Deionized water<6L/h
|
|
Humidity
|
40~85%
|
|
Size&Weight
|
External Size(W*D*H)
|
873*585*500mm
|
|
Net Weight
|
80kg
|
|
Package Size(W*D*H)
|
1130*735*1040mm
|
|
Gross Weight
|
125kg
|