General Parameters
|
Item |
Specifications |
| Power Supply |
AC220V, 50/60Hz(Standard) 1320VA; AC110V, 60Hz(Optional) |
| Size(L*W*H) |
950*645*(1250~1450)mm |
| Maximum Heater Output Power(220VAC) |
Air Chamber Heater: 330W
Water Tank Heater: 155W |
| Auxiliary Network Power Output |
AC220V/50Hz, both maximum permissible currents are 220V 1.5A |
| Heater Indication |
0~100%, 10% increasing amount |
| Temperature Control Mode |
Bed temperature mode; The skin temperature mode |
| Bed Temperature Control Temperature Range |
25.0~37.0℃;
37.1~39.0℃(Span mode) |
| Skin Temperature Control Range |
34.0~37.0℃; 37.1~38.0℃(Span mode) |
| Skin Temperature Sensor Display Range |
5.0~65.0℃ |
| Heating Time |
≤50min |
| Temperature Variability |
≤0.5℃ |
| Temperature Uniformity(Mattress in horizontal position) |
≤0.8℃ |
| Temperature Uniformity(Mattress in tilted position) |
≤1.0℃ |
| Difference Between Average Incubator Temperature and Control Temperature |
≤1.5℃ |
| Skin Temperature Sensor Accuracy |
±0.3℃ |
|
Humidity Control
|
The running time of the tank after it is filled with distilled water |
Humidity below 70%RH, at least 24h |
| Water Tank Capacity |
1500ml |
| Humidity Display Range |
0%RH~99%RH |
| Humidity Control Range |
CAH%RH~90%RH(1% per adjustment) |
| Humidity Control Accuracy |
±10% |
|
Oxygen Concentration Control
|
Oxygen Concentration Display Range |
0%~99% |
| Oxygen Concentration Display Resolution |
1% |
| Oxygen Concentration Control Accuracy |
±5% oxygen volume concentration |
| Oxygen Concentration Setting Range |
21%~60%(1% per adjustment) |
| Oxygen Sensor Service Life |
The maximum use time is 10000h at 100% oxygen concentration |
|
Transport Storage
|
Temperature |
-20℃~+55℃ |
| Atmospheric Pressure Range |
500hPa~1060hPa |
| Relative Humidity |
≤93%RH |
| Packing Size/Gross Weight |
1150*800*1600mm
190kg |
|
The Work Environment
|
Relative Humidity |
30%RH~75%RH |
| The Environment Temperature |
+20℃~+30℃ |
| Pressure Range |
700hPa~1060hPa |
| Service Altitude |
≤2000m |
| Over Voltage Category |
II |
| Pollution Levels |
2 |
| Ambient Air Velocity |
≤0.3m/s |
|
Other Indicators
|
Environmental Noise In Incubator Baby Compartment |
No more than 55dB(A) at stable temperature |
| Carbon Dioxide Concentration In The Infant Compartment |
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization |
| The velocity of air flow over a baby’s mattress |
≤0.35m/s |
| General
Alarm |
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm
|
| The System Alarm |
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure
|